Huizhou Yang
← All projects

Cementless Fixation in Partial Knee Replacement

2024–2025 · Unicompartmental knee arthroplasty

Partial knee implant and bone models, followed by tibial and femoral interface micromotion maps
Cementless partial knee replacement models, bone stiffness maps, and predicted implant–bone micromotion from the original Projects overview.

Background

Cementless partial knee implants rely on initial fixation while bone grows into the porous surface. Implant placement, patient bone properties, and design features can all affect movement at the interface.

Methods

Three finite element knee models simulated gait, deep knee bending, and stair descent. The study varied component alignment, bone stiffness, anterior–posterior translation, and fixation features, then compared interface micromotion and the porous area favorable for bone ingrowth.

Key Findings

  • Peak tibial tray micromotion was higher than femoral component micromotion and consistently occurred along the lateral aspect of the tibial tray.
  • In the modeled medial partial knee replacements, moving the tray 1 mm medially or the femoral component 1 mm laterally reduced tibial micromotion by 19.3% and 26.3%, respectively.
  • A 5 mm posterior shift of the femur increased tray–bone micromotion by 35.8%.

Significance

The results identify component positioning and load transfer as factors to consider when evaluating cementless fixation in partial knee replacement.